It’s useful to review a little bit about why we do phototherapy. All neonates have some elevation in bilirubin, due to a variety of physiologic factors. Bilirubin comes from the breakdown of the heme in red blood cells, which creates unconjugated bilirubin. Unconjugated bilirubin is fat soluble and can cross the blood brain barrier, and if present at high levels can cause acute and chronic neurologic sequelae. Bilirubin must be made water soluble before it can be effectively excreted in the bile and urine. Unconjugated bilirubin can be conjugated in the liver, or we can photo-isomerize it in the skin, either of which allow it to be eliminated.
Photo-isomerization happens most effectively when the light is around 460-470nm in wavelength, which is a blue color. Anything in the 420-500nm range works to a degree. Phototherapy irradiance is measured in units of microwatts per square centimeter of exposed skin per nanometer of blue light spectrum. Sunlight is typically in the range of 5-10 μW/cm2/nm; the issue with sunlight is that it also contains light in the ultraviolet (<400nm) wavelengths, and so must be filtered. “Standard” phototherapy should be around 8-10, whereas “intensive” phototherapy is defined as >30 μW/cm2/nm. In general, for term children, there is no reason not to do intensive phototherapy, but very small infants (<1000g) should probably be limited to < 25 μW/cm2/nm as higher levels have the potential to cause neurologic injury (although there is still no consensus on the safe or appropriate level for premature infants).
My personal experience is that most phototherapy units in circulation in low and middle income countries do not even hit standard irradiance thresholds, let alone intensive.
Irradiance is measured in μW/cm2/nm. This is the amount of power in micro watts, per square centemeter, divided by the number of nm of blue light spectrum measured by the device. Some devices will report in μW/cm2/nm; other devices will report in μW/cm2, which then needs to be divided by the nm of spectrum measured. So, for instance, if the device is reading 744μW/cm2, and the spectrum measured is 422-499nm, then the irradiance is 744/(499-422) = 744/77 = 9.7 μW/cm2/nm.
While you can use halogens, white compact fluorescents, or blue compact fluorescents, LEDs are the best choice, in my opinion. All of these can make an effective phototherapy unit, but LEDs are by far the least expensive and require the least maintenance. To put some numbers to it, the compact fluorescent units that my mission hospital had been using were built with 6 tubes, each costing USD $30. When you add in the lamp holders and the carpentry, the units cost around $230 to build. These would reach an irradiance of 10-14 μW/cm2/nm with fresh bulbs, but would rapidly degrade with heavy use, and the bulbs would need to be replaced every 3-6 months. By contrast, an LED unit costs about $35 to build, and the power supplies seem to go bad every 1-2 years, which cost about $15 to replace, although typically they can be fixed by swapping out the capacitors.
LEDs can be powered in one of two ways. The more efficient way is to use a constant current DC power supply (also called a constant current driver); this will vary the voltage within a range to maintain the target amperage through the LEDs. Constant voltage power supplies, by contrast, supply a fixed voltage but require a current limiting device on each LED to prevent overcurrent. This typically is in the form of a paired resistor, which reduces efficiency and generates more heat. While a poorer technical choice for these reasons, constant voltage power supplies are much more available and are usually less expensive, which is why our design uses them. LED bulbs that insert into a socket have the driver (typically constant current) built into the base.
Now for the LEDs themselves – thankfully, commercial blue LEDs are typically of the iridium-gallium type which produces light peaking in the 450-470nm wavelength range, which is just where we want it. “White” LEDs are made by combining red, blue, and green LEDs – so they will work, as they contain blue light, but are less efficient for our purposes. The same is true for color changing LEDs. Due to cost and availability, we use LEDs strips for most of our phototherapy units. These usually come in 5 meter rolls with 300 LEDs per roll. Each LED has a paired resistor so they work with constant voltage drivers – the standard is 12V. You will usually see them listed as SMD5050 or SMD3528 or some other number. SMD stands for “surface mounted device,” and the numbers following are the dimensions of the LED in 0.1mm. The ones I usually use are 5050, which have LEDs that are 5mm x 5mm, as these put out more light per LED, but any will do – you’ll just need to adjust the number of strips. Look for “blue,” not “icy blue.” Some manufacturers will list the wavelength but most won’t.
For 300 LED SMD5050 rolls, each 5 meter roll usually uses around 72W. So you’ll need a 12 volt DC driver that will supply 72W/12V = 6A per roll. The units we build use 2 rolls, so I typically use a 180W 15A power supply.
LED Strip Lights
Power Supply
Miscellaneous Electrical Supplies (optional thermal fuse also pictured)
Soldering supplies and electrical tools
Wood for frame (wire mesh and screws not pictured)
14G Wire
Woodworking tools (saw not pictured)
Irradiance Meter, optional but highly recommended
You’ll need the following. These may be available locally; if not, I have had good results ordering from aliexpress.com.
Two 5 meter long (300 LED) 12V SMD5050 strips (usually less than $5 each); if using LEDs smaller than 5050 you’ll need more.
One 180W 12V 15A power supply (usually around $15-20). Any DC power supply will do, and it’s OK if the amperage/wattage is higher, it just needs to be 12V. If you can only find smaller ones that’s OK too, you’ll just need to combine them.
Basic soldering supplies (less than $1 of materials). The electrical work is well within the skillset of an electrical appliance repairman, so if you don’t solder yourself often and don’t feel like learning from YouTube then I’d recommend hiring someone for the soldering work.
A cord and plug ($1-2), some 12-14 gauge (1.5-2mm) wire ($1-2), and a switch (< $1) to toggle between high intensity and lower intensity. If your plug is not fused I would recommend adding a fuse to the AC side of the circuit.
Wood, screws, and paint or polyurethane ($5).
Optional but recommended: surge protector and thermal fuse (I use 77C but the exact number is not critical, I would recommend in the 70-100C range)
Optional but strongly recommended: a blue light irradiance meter ($250). I ordered from a Solarmeter brand unit from Amazon.com and had it shipped to my country. I know it is expensive but I think it is a worthwhile investment. I’ve tried using the luxmeter built into my phone but could not get consistent results. I think you might be able to make a very inexpensive meter with a photovoltaic cell and a multimeter but I haven’t had time to experiment to see how reproducible it is.
Total cost is around $35, time required to build a unit if you’ve never build one before is about 2 hours.
The exact composition will depend on local material availability and expertise; you could use solid wood, plywood, block board, or even metal. We use plywood for most of it and blockboard for the sides of the power supply enclosure. Make sure that the top is not too thick or else you will have more problems with heat buildup, although you could solve this by adding a 12V computer case fan. Our design uses a enclosure around the power supply to provide some protection from accidental shocks.
The dimensions of your phototherapy units will depend how your incubators and bassinets are constructed. The tops of our incubators are plexiglass sheets which will accommodate a phototherapy unit 36.5 x 74.5 cm. I had a local carpenter construct the box out of 6mm plywood, with blockboard around the power supply. The main box is 36.5 x 74.5 x 7cm deep, with another box 36.5 x 15cm x 7cm deep atop it for the power supply. The inside is painted white to maximize reflectance, and the outside is polyurethaned.
Our units use two 5 meter strips of 5050 LEDs, each of which has 60 LEDs per meter. There are places where you can cut every 5cm marked with a scissors icon. I cut them into 15 strips of 65cm and space them evenly 1cm apart (there is one strip that has a piece from each roll).
Affix the cut strips to the inside of the case. The strips have an adhesive backing that is of varying quality manufacturer to manufacturer and will sometimes need to be reinforced with glue. You can remove the pre-soldered wires at the ends of the roll. If you are using 65cm strips, there will be one strip made of two pieces from the end of one roll and the start of the next - align the cut copper ovals next to each other and we will solder together in the next step.
Each strip has a positive and a negative end - usually the negative side is marked with a "-". Make sure that you put all the strips in the same orientation (eg all the negatives up).
To complete the circuit, all the positive rails must be connected to the positive terminal of the power supply, and all the negative rails to the negative terminal. The connection points are the little copper ovals. It doesn't matter where on the strip is connected, as long as one of the ovals on that strip is connected back to the appropriate terminal.
To allow us to reduce the intensity for preemies, there is a slight variation on the positive rails. Nine of the positive rails connect back to the power supply directly. The other six have a switch in between that can disconnect and them from the supply, allowing them to turn off.
There are any number of ways to do the soldering. If you have never soldered before, that's OK - this is a skill that you can easily pick up with a YouTube video and a few minutes of practice. Or, if you can find an appliance repair person, this is well within their skillset.
The way I find quickest is to take a strip of wire and remove about 6cm of insulation completely. Then cut 14 circumferential slits in the insulation about every 2cm, creating short segments of insulation which you can slide up and down the wire. Now you can position the insulation so that it covers the copper oval which you are not soldering to (see picture).
Now it's time to connect all the wires. Obviously, do this while everything is powered off. The dangerous parts are the AC connections - the 12V DC connections shouldn't shock you even if you touch them while powered, although I wouldn't make a practice of it.
Connect the wire going to all the negative side of all the strips to the negative terminal of your power supply.
Connect the wire going to the positive connection on approximately 2/3 of your strips (in my setup, 9 of 15 strips) directly to the positive terminal of your power supply.
Connect the wire going to approximately 1/3 of your strips (in my setup, 6 of the 15 strips) to a switch and then to the positive terminal of your power supply. This will allow you to turn on and off these strips to lower the intensity as needed.
If you are using a 12V fan, which may improve longevity of the power supply, attach it to the positive and negative terminals of the power supply. Note that you can have multiple connections to the same terminal if needed.
Connect the live, neutral, and ground coming from the mains AC to the appropriate terminals. If using a thermal fuse, connect it between the live wire and the live terminal. If you plug does not have a fuse built in, connect a regular fuse in as well. The thermal fuse should blow if the device has a short and becomes very hot, and the regular fuse should also blow if there is a large short. Thankfully, I've never had either of these happen.
I add a screen mesh over the power supply for safety to keep people from accidentally touching the high voltage AC wires.
Turn on the power supply (as an aside, I strongly recommend using a surge protector if your mains power is not stable) and make sure everything lights up as expected. Occasionally you'll get a bad strip where there is a discontinuity somewhere in the embedded positive or negative copper strips. In that case, you can test by holding a wire between the copper oval that starts the unlit section and the copper oval of a lit adjacent strip. Make sure to do positive to positive and negative to negative. If that causes the LEDs to light up, solder that wire in place.
When measuring the unit, make sure to do it "in situ" - at the same distance and with the same surroundings where it will be used. See the next section for more information on measuring.
If it is underpowered, first check to see if you have a bad power supply, either by swapping out with a known good supply or measuring the current and voltage with a multimeter. If that is working correctly, next try optimizing the distance to the infant and adding in white reflectors if possible. If your irradiance measurements are still low, then you can add in additional strips of LEDs. You may have to increase the power your your supply depending on how many strips you add.
I have found that even using different manufactures of power supplies and LEDs, the readings are pretty consistent unit to unit. This means that if you can at least borrow an irradiance meter and show one unit is giving appropriate irradance in your typical setup, others built to the same specifications probably will as well.
The reading to the left is from this unit placed over an incubator, corresponding to an irradiance of 39 μW/cm2/nm. Measured again with the "preemie switch" and about 5cm more distant from the lights (to account for the smaller girth of a premature infant) , the measured irradiance was 18 μW/cm2/nm. Both within targets - success!
If you have a irradiance meter, which is an investment I strongly recommend if you can afford it, you should regularly test your units to make sure they are continuing to hit target levels. If you do not have a meter, I have some general conservative recommendations about replacement schedule, but these are more theoretical than tested.
We test our units monthly, and as needed if they appear less "bright" to the eye. To test the device, place the irradiance meter at the distance the baby's skin will be from the phototherapy unit (keeping in mind that premature babies will be further on average) and move it around the testing field to get a feel for what the average reading will be. Test "in-situ," using the incubators or bassinets that are used in your setting. In our setup the range is about 3 μW/cm2/nm difference between the brightest and dimmest locations in the incubator. I tend to use the low average value, as in general I would rather be overpowered than underpowered. We have been using the units for several years now and the only component that has needed repair or replacement has been the power supplies, but if the power starts to decrease it would be easy to add or replace LED strips.
If you do not have an irradiance meter, I would check the current output of the power supply monthly with a multimeter, and replace it if it drops by >10%. I would also routinely replace the LEDs every 2 years of typical use. This is very conservative: LEDs are probably good for around 10,000-20,000 hours; if we assume around 12h use on average per day this would mean we replace well before the 10,000 hour mark.
Here is the maintenance sheet we use:
Note that our irradiance meter measures in μW/cm2, over a 77nm spectrum, and also has a maximum reading of 1999μW/cm (which corresponds to 26μW/cm2/nm). We affixed a mesh over the sensor which reduces the reading to 40% of the unfiltered value, meaning we divide the reading by 30 (40% of 77) to get the reading, and expanding our detectable range to around 65 μW/cm2/nm. See below for instructions on how to modify your meter if you have similar problems.
You will need to modify the instructions based on how your meter provides readings.
As mentioned above, our irradiance meter was not able to read high enough to measure our intensive units, pegging out at 26μW/cm2/nm. To modify it, I used a nylon tea strainer mesh and doubled it up, forming it into a dome shape. This is taped to a cardboard sleeve which fits snugly around the meter. The mesh is then slipped over the sensor. Make sure that the mesh is always a fixed distance from the sensor or else it may make the attenuation less consistent. I tested the meter in a number of different settings with and without the mesh to determine the percent attenuation, and it was consistently measuring between 39-41% of the unfiltered reading.
Close up of the mesh filter
The irradiance meter with the mesh in place over the sensor
Thank you to Dr Tina Slusher and Dr Hendrik Vreman for lending their expertise (and spectrometer).
Soli Deo gloria.